Containment Barrier Isolator (CBI) | Safe Glove Change | Esco Pharma

Описание к видео Containment Barrier Isolator (CBI) | Safe Glove Change | Esco Pharma

Esco shares these steps for the CBI Safe glove change:

1. Remove the inner and outer O-rings.
2. Carefully roll the gloves from the innermost to the outermost groove of the cuff ring.
3. Put a new glove over the old glove up to the innermost groove of the cuff ring.
4. Install the O-ring to the innermost groove.
5. Detach the old glove (inside) from the cuff ring.
6. Put back the O-ring on the outermost groove.
7. Dispose of the old glove to a safe change bag and send it to the tight pass chamber.
8. Unscrew and remove the glove port cover.
9. Remove the O-ring and carefully lower down the glove to the last groove.
10. Put on the new glove and remove the old glove (inside) through the adjacent glove port.
11. Put back the O-ring and the glove part cover.
12. Dispose of the old glove to a safe change bag and send it to the right pass chamber.

-------------------------------------------------------------------------------------------

The Esco Lifesciences Group, headquartered in Singapore, is building a synergistic ecosystem of life science tools, diagnostics, therapeutics, and technologies to help improve human lives.

To learn more about our world-class products and services, visit our website: https://www.escolifesciences.com/

Subscribe to our channel: http://bit.ly/EscoLifeSciences

Follow us on our social media accounts to get the latest news and updates.