How To Insert the i-Port Advance® Injection Port

Описание к видео How To Insert the i-Port Advance® Injection Port

In this video, you'll receive step-by-step instructions on how to apply, use, and remove i-Port Advance injection port. Begin by washing your hands thoroughly with soap and water. Then find an application site that works for you. Application sites include the abdomen, upper thighs, back of arms, hips and upper buttocks. Be sure to avoid areas that rub against clothing such as a belt or the waistband of your pant line.

Next, clean the application site you've chosen by swabbing your skin with rubbing alcohol. Be sure to let the skin surface dry completely before applying the i-Port Advance injection port. Pull down on the red tab located on the side of the injection port and remove the plastic seal wrapped around the device. Next, remove the sterile paper from the bottom of the device by gently pulling back on the tab. Turn the device over, and locate the three raised marks on either side of the lid. Squeeze the raised marks to remove lid. Next, remove the paper backing from the adhesive pad of the i-Port Advance injection port by gently pulling up on the white tab. Remove the needle guard by gently twisting to loosen it and pulling it off. Turn the insertion device over, and with the insertion needle facing away from you, hold the lined ridges located on either side of the device. With the other hand, pull on the center of the insertion device until you hear it click. You are now ready to apply the device.

To apply, hold the insertion device against your skin over the site you have prepared. Locate the round indentations on either side of the device. Once your fingers are in position, squeeze simultaneously to insert the injection port. Press down on the center of the insertion device to secure the i-Port Advance injection port to the skin and then gently pull the insertion device straight out. Smooth down the adhesive to secure it to your skin. To discard the insertion device, replace the plastic lid and dispose of it in your sharps container. The i-Port Advance injection port is now ready to be used.

Remember, do not use needles longer than 5/16th inch or 8 millimeters, or smaller than 3/16th inch or 5 millimeters. Also, do not use needles thicker than 28 gauge. Swab the septum or top of the i-Port Advance injection port. To begin injection, insert the needle of the syringe or pen fully into the septum and slowly push down the plunger. Medication will be delivered to the tissue under your skin as shown in this graphic. If you are using a pen, refer to the manufacturer of your medication to determine how long to wait to administer your medication. Hold the i-Port Advance injection port down with your free hand, and slowly remove the syringe or pen. For individuals with diabetes that require both long acting and rapid acting insulin, it is suggested to inject the rapid acting insulin first, wait 60 minutes and then inject the long acting insulin. You can wear the i-Port Advance injection port for up to 72-hours before changing injection sites and you can continue participating in all normal daily activities such as bathing, sleeping, and exercising. To remove the i-Port Advance injection port, grab the edge of the adhesive and pull off. Dispose of the used i-Port Advance injection port according to your state's environmental health and safety regulations. For more information, please review the user guide.

Комментарии

Информация по комментариям в разработке