easy operate method of the fiber splicer Arc Calibration for AI-7,AI-7C,AI-8,AI-8C,AI-9

Описание к видео easy operate method of the fiber splicer Arc Calibration for AI-7,AI-7C,AI-8,AI-8C,AI-9

Arc Calibration

Situation: when you find the fiber fusion voltage is too low or when fusion processing terminal not fully fusion, you can have a try of this operation.

Steps:
1、connect app and choose the “discharge correction” button.
2、See the machine from “auto mode ” to “Discharge Correct mode”.
3、Put the fiber processing and then put in the machine.
4、After close the cover, it comes auto Discharge Correct.
5、Listen to the voice leader, if it remind you “please cleave the fiber and continue the fiber arc calibration”, it means the voltage is not enough and we need do the job 3 and 4 again. Until it remind you “arc calibration succeed”. then you can normally use it.